TOPAZ Dual UV Detector

Two wavelength in one cell 200 up to 320 nm

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
73E000000

TOPAZ Dual UV Detector

is a new type of UV detector, which allows measuring absorbance of two wavelengths simultaneously in one cell. This unit is used in liquid chromatography to verify analyzed samples by means of a second wavelength or in situations when some peaks absorb on different wavelengths. It is possible to use the detector in analytical, preparative or flash applications by using different cells. Noise level at 254 nm is ± 1 x 10-4 AU with an empty cell.

The unit’s DAD (diode array detector) design offers many advantages:

  • A wavelength setting may be changed in steps of 1nm
  • The unit is easy for operating, reference and sample signal are available for detector diagnostics
  • Output signal is it available in digital and analogue form on the output connector
  • Work hours of lamp are counted using build in counter.
  • Detector cell is easy replaceable from the side of the detector
 

Important notice: Output signal of the detector is directly proportional to the optical path of the cell. Since the optical paths of cells may be changed, noise and drift levels are referred to the optical path of 10 mm, corresponding to the AU definition.

Used cells

Preparative cell PLCC 14 (supplied with unit)

volume/optical path(adjustable) 45 µl / 0.1 mm, 55 µl / 1.4 mm, 70 µl / 2.4 mm
connecting tubing with OD = 1/16" and 1/8", thread 1“-28

Analytical cell HPLC 08 (optionally)

Volume / optical path 10 µl / 5 mm
Diameters of capillaries OD x ID input 1/16" x 0,2 mm; output 1/16" x 0,5 mm

 

Accuracy of adjustment ± 1 nm
Typical spectral half-width 8 nm
Reproducibility ± 0.5 nm
Light source Deuterium discharge lamp
Output for integrator 1 V/AU
Time constant (T90) 1 s
Communication RS232
Power supply 230 V ± 10%, 50 Hz , 115/100 V± 10%, 50-60 Hz
Power input 75 VA
Dimensions W x H x D 220 x 178 x 435 mm
Weight 12 kg
Noise level at empty cell (254 nm, TC 1 s) ± 1 x 10-4 AU
Drift at empty cell (254 nm after 1 h) 5 x 10-4 AU/hr
Materials in contact with mobile phase PTFE; fused silica, stainless steel, Vespel, PEEK