is small simple unit for use in analytical liquid chromatography applications based on measurement of absorption of UV radiation of 254nm. The unit is easy for operating, reference and sample signal is available for detector diagnostics and lamp live reading.
The detector is equipped with two interference filters 254 nm to reduce the others spectral lines of mercury lamp.
Very interesting are the small dimensions of unit.
There is possible to use cell:
Analytical cell HPLC 08
| Wavelength | 254 nm |
| Light source | low pressure mercury lamp |
| Time constant | 1.0 sec. |
| Noise level at empty cell | ± 2,5 x 10-5 AU |
| Drift at empty cell (after 1 hour) | 5 x 10-4 AU/hr |
| Absorbance output | 1V/au |
| Power supply | 230 V ± 10%, 50 Hz or 115/100 V± 10%, 50-60 Hz |
| Power input | 34 VA |
| Dimensions (W x H x D) | 153 x 112 x 210 mm |
| Weight | 3kg |
Important notice: Output signal of the detector is directly proportional to the optical path of the cell. Since the optical paths of cells may be changed, noise and drift levels are referred to the optical path of 10 mm, corresponding to the AU definition.
HPLC 08 Standard measuring analytical cell supplied with the unit
| volume / optical path | 10 ul / 5 mm |
| External diameter of capillaries | 1,6mm (1/16") |
| Internal diameter of capillaries input/output | 0.2 / 0.5 mm |