Column oven is important to achieve better temperature stability of HPLC columns by analysis.
Using it we get:
Range of temperature regulation is from ambient temperature up to 99°C. Actual oven temperature is shown on display. Temperature stability is ± 0.1°C.
Internal temperature controlled area is 21 x 23 x 205 mm so that the oven can easily accommodate a 150 mm long column.

The unit is great at its small dimensions. Heated part is separated from controlling unit which allow to install it easy into chromatography configuration. With regard to its small weight is it possible to easy hang it onto column.
Closing using fastening system make easy to install column into heating unit.
Required temperature setting is simple by means of arrow buttons and storing. After unit turning off remembers the oven its last used temperature so there is not need to enter it again by repeated analysis.
An independent safety circuit turns the heater off and transmits an acoustic signal, when set temperature is exceeded by 10°C.
| Temperature range | Ambient to +99°C |
| Temperature setting | in 0,1°C steps |
| Accuracy of temperature setting | ± 0,5°C |
| Temperature stability | ± 0.1°C |
| Heating rate | 5°C/min |
| Internal temperature controlled area | 21 x 23 x 205 mm |
| Used column dimensions | max. 150 mm; diameter max. 20 mm |
| Supply voltage | 100/115/230 V ± 10%, 50/60 Hz |
| Power input | 50 VA |
| Dimensions of heated unit (W x H x D) | 44 x 47 x 232 mm |
| Dimensions of control unit (W x H x D) | 100 x 155 x 190 |
| Weight of heated unit | 0,62 kg |
| Weight of control unit | 2,5 kg |