For detectors: LCD 2070 , LCD 2071.2, LCD 2071.3, LCD 2073A, OPAL, TOPAZ and FLASH 06.
HPLC 08 Analytical Flow-cell is used in the LCD 2070, LCD 2071.2, TOPAZ Dual UV, LCD 2073 and FLASH 06 detectors. Cell volume is 10 µl and its optical path is 5 mm. A hydrodynamic resistor is supplied with the cell to eliminate formation of bubbles. It is connected on the output capillaries. Internal diameter of capillaries input is 0.2 mm.
| Volume / optical path | 10 ul / 5 mm |
| External diameter of capillaries | 1/16" |
| Internal diameter of capillaries | input/output 0.2 / 0.5 mm |
| Maximum operating pressure | 1 MPa |
| Maximum flow-rate | 10 ml/min. |
| Casing dimensions | Diameter 36 mm, Length 23 mm |
| Weight | 70 g |
| Material | Vespel, stainless steel, fused silica |