HPLC 08 EX Flow-cell is analytical cell equipped with SMA connectors for fiber optics connection. Cell volume is 10 µl and its optical path is 5 mm. A hydrodynamic resistor is supplied with the cell to eliminate formation of bubbles. It is connected on the output capillary. Internal diameter of input capillary is 0.2 mm.
The cell is assembled outside of detector and it is connected with unit using optical fiber cable. Therefore it is used in explosion or contamination hazard applications.
UV Detectors for this cell signal evaluation have to be assembled with SMA output and input connectors. From ECOM production they are detectors with optical connectors.
| Internal diameter of capillaries - input/output | 0.2 / 0.5 mm |
| External diameter of capillaries | 1/16" |
| Volume / optical path | 10 µl / 5 mm |
| Maximum operating pressure | 1 MPa |
| Maximum flow-rate | 10 ml/min. |
| Casing dimensions | Diameter 36 mm, Length 23 mm |
| Weight | 150 g |
| Material | Vespel, stainless steel, fused silica |