Four wavelengths simultaneously,
scan of whole spectrum 200 - 400 nm
Flash 06 DAD 400 is an OEM build-in type of UV diode array detector, which allows measuring absorbance of four wavelengths simultaneously in one cell just as measuring of whole spectrum (scan). This unit is used in liquid chromatography to verify analyzed samples by means of four wavelengths or in situations when some peaks absorb on different wavelengths. It is possible to use the detector in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose. Noise level at 254 nm is ± 5 x 10-5 AU with a test cell.
The unit is designed as build in unit with 24-36V power supply and RS232 controlling. On request can be sold in classical case with 90-264 V powering.
The unit’s DAD (diode array detector) design offers many advantages:
| Wavelength range | 200 - 400 nm (256 elements on CCD) |
| Typical spectral half-width | 8 nm |
| Accuracy of adjustment | ± 1 nm |
| Reproducibility | ± 0.5 nm |
| Light source | Deuterium discharge lamp |
| Noise level at test cell (254 nm, TC 0.75 s) | ± 5 x 10-5 AU |
| Drift at test cell (254 nm after 1 hr) | 1 x 10-3 AU/hr |
| Materials in contact with mobile phase | PTFE; fused silica, stainless steel |
| Time constant (T90) | 0.5 s; 0.75 s; 1 s or 2 s (T63 0.3s;0.4s;0.6s;1s) |
| Output for integrator | 1 V/AU (in digital form only) |
| Interface | RS 232 |
| Power supply | 24-36 V DC |
| Power input | 48 W |
| Dimensions (W x H x D) | 131 x 167 x 386 mm |
| Weight | 4.3 kg |
| volume/optical path(adjustable) | 45 μl / 0.3 mm (basic) 55 μl / 1.4 mm 70 μl / 2.4 mm |
| connecting | tubing with OD = 1/8", thread 1/4"-28 |
| maximal flow rate | 500 ml/min |