TOYDAD 400 Scaning UV Detector

Scanning 200-400

200 - 400 nm Scanning  PDA detector  measures at four wavelengths simultaneously or sends scan with speed 5 Hz which allows to create 3D picture.

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
TOY0300X

More wavelengths or scan are used in liquid chromatography to verify purity of analyzed samples or in situations when some peaks absorb on different wavelengths. 

Most important features of TOYDAD detectors:
• small dimensions
built-in lamp working hours counter
easy replacement of flow cell
measuring at more wavelengths simultaneously
sophisticated diagnostic software

  

PREPARATIVE CELL PLCC 15 (SUPPLIED WITH UNIT)

volume/optical path (adjustable) 45 μl / 0.3 mm; 55 μl / 1.4 mm; 70 μl / 2.4 mm
Cell connecting tubing with OD = 1/8", thread 1/4"-28
Number of channels (Signals) 4
Scan 200 - 400 nm, 5 Hz, step 1 nm
Number of CCD elements 128
Accuracy of adjustment ± 1 nm
Typical spectral half-width 8 nm
Reproducibility ±0.5 nm
Light source Deuterium lamp
Noise (Test cell, 254nm, TC 2s, 10 Hz) ± 5 x 10-5 AU
Drift (Test cell, 254nm) 1 x 10-3 AU/hr
Time constant 0.5 s, 0.75 s, 1.0 s, 2.0 s, 4.0 s, 8.0 s, 16.0 s, 0.2 s, 0.1 s
Wetted materials FEP; fused silica, stainless steel
Communication RS232
Power supply 24 -36 V DC
Power input 48 W
Dimensions W x H x D 131 x 167 x 386 mm (5.15 x 6.57 x 15.20 in)
Weight 3.20 kg (7.06 lb)
Optical cables no
Filename Type