Two wavelength (200 up to 320 nm) in one external cell
is a UV detector, which allows measuring absorbance of two wavelengths simultaneously in one cell

connected using fibre optic cable with SMA connectors.

This unit is used in liquid chromatography to verify analyzed samples by means of a second wavelength or in situations when some peaks absorb on different wavelengths.
Noise level at 254 nm is ± 1 x 10-4 AU with an empty cell.
The unit’s DAD (diode array detector) design offers many advantages:
| Wavelength range | 200 - 320 nm |
| Spectral half-width | 8 nm |
| Accuracy of adjustment | ± 1 nm |
| Reproducibility | ± 0.5 nm |
| Light source | Deuterium discharge lamp |
| Noise level at empty cell (240 nm, TC 1 s) | ± 1 x 10-4 AU |
| Drift at empty cell (240 nm after 1 h) | 5 x 10-4 AU/hr |
| Output and input optical connector | SMA |
| Time constant | 1 s |
| Output for integrator | 1 V/AU |
| Interface | RS 232 |
| Power supply | 230 V ± 10%, 50 Hz , 115/100 V± 10%, 50-60 Hz |
| Power input | 75 VA |
| Dimensions (W x H x D) | 220 x 178 x 435 mm |
| Weight | 12 kg |
Preparative cell PLCC 20 (optionally)
| Optical path – Basic continuously changeable | 0.1 to 1.5 mm |
| Discontinuous changes of basic optical path | +1 and +2 mm |
| Connecting thread of inlet nipple | NPT 1/4”-18 |
| Inlet nipple I.D. | 1/4” (6,35 mm) |
| Internal cell diameter | 8 mm |
| Maximum flow rate | 15 l/min by pressure 0,5 MPa |