UV-VIS Detector with fixed wavelength and optical connectors. Interference filter determining wavelength is ordered separately.
is a detector with an interference filter. By means of this filter can be this unit adjusted at one fixed wavelength in range of 200 – 600 nm. Noise level at 254 nm is ± 0.5 x 10-5 AU with an empty cell. (Filter is not include price)
Cell is installed out of unit and it is connected to the unit using fiber optic cable with SMA connectors.

It is possible connect the detector to a cell for ITP or with cell PLCC 20 use it for preparative or flash chromatography.

The unit is easy to operate, reference and sample signals are available for detector diagnostics as well as information on lamp operating hours.
Output signal is available in both digital and analogue form on the output connector. It is possible to control the unit using the keypad or an RS232 interface.
The detector is equipped with a high standard deuterium lamp or tungsten lamp in a special socket that enables easy changing.
Cell, interference filter and fibre optic cable are not include.
| Wavelength range | 200 - 600 nm |
| Spectral half-width | depending of used filter |
| Light source | Deuterium discharge or tungsten lamp |
| Noise level at empty cell (254 nm, TC 1 s) | ± 0,5 x 10-5 AU |
| Drift at empty cell (254 nm after 1 h) | 2x 10-4 AU/hr |
| Output and input optical connector | SMA |
| Time constant | Fast, middle and slow |
| Output for integrator | 1 V/AU |
| Interface | RS 232 |
| Power supply | 100, 115 and 230V, 50/60 Hz |
| Power input | 80 VA |
| Dimensions (W x H x D) | 220 x 170 x 450 mm |
| Weight | 8 kg |
| Wavelength (nm) | Spectral half width (nm) | P/N |
| 200 | 12 | 42444 |
| 214 | 12 | 42436 |
| 228 | 12 | 42443 |
| 254 | 12 | 42435 |
| 280 | 10 | 42437 |
| 313 | 10 | 42440 |
| 297 | 10 | 42442 |
| 340 | 10 | 42438 |
| 460 | 12 | 42441 |
| 560 | 10 | 42439 |
Other wavelength filters at request
Preparative cell PLCC 20 (optionally)
| Optical path – Basic continuously changeable | 0.1 to 1.5 mm |
| Discontinuous changes of basic optical path | +1 and +2 mm |
| Connecting thread of inlet nipple | NPT 1/4”-18 |
| Inlet nipple I.D. | 1/4” (6,35 mm) |
| Internal cell diameter | mm |
| Maximum flow rate | 15 l/min by pressure 0,5 MPa |