One fix wavelength given by chosen interference filter
in range 200-600 mn
FLASH 06 SINGLE is detector with an interference filter. Detector can be adjusted using interference filter at one fixed wavelength in range of 200–600 nm. Noise level at 254 nm is ± 2 x 10 -5 AU with a test cell. (Filter 254 nm include price, other filters on request). High standard deuterium lamp in a special socket enables easy changing. This unit is used in liquid chromatography in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose.
The unit is designed as stand alone with 90–264 V power supply and RS232 controlling. It is derived from OEM build-in unit with powering 24-36 V DC.
The unit offers many advantages:
| Wavelength range-according used interference filter | 200 - 600 nm (standard installed filter 254 nm) |
| Typical spectral half-width | According used interference filter |
| Light source | Deuterium discharge lamp |
| Noise level at test cell (254 nm, TC 1.5 s) | ± 2 x 10-5 AU |
| Drift at test cell (254 nm after 1 h) | 1 x 10-3 AU/hr |
| Materials in contact with mobile phase | PTFE; fused silica, stainless steel |
| Time constant (T63) | 0 s; 0.3 s; 0.6 s; 1.5 s |
| Output for integrator | 1 V/AU (in digital form only) |
| Interface | RS232 |
| Power supply | 90–264 V |
| Power input | 90 VA |
| Dimensions (W x H x D) | 220 x 170 x 450 mm |
| Weight | 7.9 kg |
| volume/optical path(adjustable) | 45 μl / 0.3 mm (basic) 55 μl / 1.4 mm 70 μl / 2.4 mm |
| connecting | tubing with OD = 1/8", thread 1/4"-28 |
| maximal flow rate | 500 ml/min |