AS96 Autosampler

New Product

Automatic autosampler for 96 vials, with or without a cooling and heating module, 40 MPa or 60 MPa

Availability
Categories
Manufacturer
-
Choose variant:
Image PN Cooling and heating module Maximum pressure Request quote
Image
With cooling and heating, 60 MPa, ASA96000
PN
ASA96000
Cooling and heating module
yes
Maximum pressure
60 MPa (8702 psi)
Actions
Image
Without cooling and heating, 60 MPa, ASA96010
PN
ASA96010
Cooling and heating module
not
Maximum pressure
60 MPa (8702 psi)
Actions
Image
Without cooling and heating, 40 MPa, ASA97000
PN
ASA97000
Cooling and heating module
not
Maximum pressure
40 MPa (5801 psi)
Actions
Image
Withcooling and heating, 40 MPa, ASA97010
PN
ASA97010
Cooling and heating module
yes
Maximum pressure
40 MPa (5801 psi)
Actions

A reliable and precise autosampler for a wide range of applications in HPLC and LC-MS with/without cooling and heating, in a compact design. 
AS96 Autosampler is a generic autosampler using state-of-the-art injection technology with fast injection and wash cycles. Efficient, multi-solvent needle wash virtually eliminates carry-over. AS96 enables excellent reagent addition and mix capabilities for derivatization, dilution, and internal standard addition. Our Pressure Assisted Sample Aspiration (PASA™) concept avoids sample-syringe contact and air bubbles and has proven its robustness in more than 30,000 autosamplers. AS96 Autosampler handles well plates and sample vials, either open or sealed. Its compact, stackable design and unrivaled performance make the AS96 the best fit for any analytical system. The autosampler can also be equipped with a sample cooling module. The Peltier module with forced air cooling provides true sample cooling down to 4 °C in the sample compartment.

Features

● Metal-free sampling with silica-coated steel needle and PEEK valve (option)
● Reagent addition and mix capabilities for derivatization, dilution, internal standard addition
● Optional 6-port solvent selection valve (SVV) for extended selection of reagents for needle wash and reagent addition (option)
● Integrated Stream Switching (ISS): optional extra high-pressure 6-port switching valve for column switching or other stream switching applications up to 18,000 Psi
● Quick-fit injection valve for fast service

ParameterValue
Injection modes Full-loop, Partial loop-fill , µL-pick-up
Injection volume Programmable from 0 µl – 9999 µL 1 µL increments
Injection precision Full-loop injection ‹ 0.3% RSD; Partial loop-fill ‹ 0.5% RSD; µL -pick-up ‹ 1.0% RSD
Sample viscosity 0.1 – 5 cP
Injections per vial/well Max 9
Syringe volume 500 µL standard; 250, 1000 and 2500 µL optional

Needle wash
Inside and outside needle wash with drying. Wash can be programmed between injections and between vials/wells.

1 solvent
5 additional wash solvents
Injection cycle time 60 seconds
Valve switching time 60 msec
Wetted parts SS316, PTFE, TEFZEL, VESPEL, glass, For Bio-kit option: PEEK and coated-steel (needle) instead of SS316
Carry-over ‹ 0.05% with standard wash
Typically ‹ 0.01% with extra wash
Sample capacity Microtiter plates: 2; Vials: 2x 48 (1.5 mL) or 2x 12 (10 mL) - optional
Maximum vial/MTP height 47 mm
Sample cooling -Factory installed option (Peltier technology) Minimum: 4°C ± 2°C Maximum: ambient temp -3°C
Syringe volume 500 µL standard; 250, 1000 and 2500 µL optional
Wetted materials SS316, PTFE, TEFZEL, VESPEL, glass, For Bio-kit option: PEEK and coated-steel (needle) instead of SS316
Loop volumes Programmable from 0 µl – 9999 µL 1 µL increments
Sample Capacity Microtiter plates: 2; Vials: 2x 48 (1.5 mL) or 2x 12 (10 mL) - optional
Injection volume Programmable from 0 µl – 9999 µL 1 µL increments
Injection Precision Full-loop injection ‹ 0.3% RSD; Partial loop-fill ‹ 0.5% RSD; µL -pick-up ‹ 1.0% RSD
Filename Type