Flash 06 Single Detector

One fix wavelength given by chosen interference filter in range 200-600 mn

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
2048000X

 FLASH 06 SINGLE is an OEM build-in detector with an interference filter. Detector can be adjusted using interference filter at one fixed wavelength in range of 200–600 nm. Noise level at 254 nm is ± 2 x 10 -5 AU with a test cell. (Filter 254 nm include priceother filters on request) High standard deuterium lamp in a special socket enables easy changing. This unit is used in liquid chromatography in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose.

The unit is designed as build in unit with 24-36V power supply and RS232 controlling. On request can be sold in classical case with 90-264 V powering.

The unit offers many advantages:

  • wavelength setting from 200 up to 600 nm according used interference filter
  • lamp work hours are counted using the built-in counter
  • the cell is easy to replace from the side of the detector
  • powering by 24-36V
  • unit is controlled by RS232 interface
 Preparative cell PLCC 15 (delivered with unit):
volume/optical path(adjustable) 45 µl / 0.3 mm (basic)
55 µl / 1.4 mm
70 µl / 2.4 mm
connecting tubing with OD = 1/8", thread 1/4"-28
maximal flow rate 500 ml/min
Typical spectral half-width According used interference filter
Light source Deuterium discharge lamp
Noise (Test cell, 254nm, TC 2s, 10 Hz) ± 2 x 10-5 AU
Drift (Test cell, 254nm) 1 x 10-3 AU/hr
Output for integrator 1 V/AU (in digital form only)
Time constant (T90) 0 s; 0.3 s; 0.6 s; 1.5 s
Communication RS232
Power supply 24-36 V DC
Power input 48 W
Dimensions W x H x D 131 x 167 x 386 mm
Weight 3.5 kg
Materials in contact with mobile phase PTFE; fused silica, stainless steel