Flash 06S Single Detector with housing

One fix wavelength given by chosen interference filter 
in range 200-600 mn

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
2048S00X

 FLASH 06 SINGLE is detector with an interference filter. Detector can be adjusted using interference filter at one fixed wavelength in range of 200–600 nm. Noise level at 254 nm is ± 2 x 10 -5 AU with a test cell. (Filter 254 nm include priceother filters on request). High standard deuterium lamp in a special socket enables easy changing. This unit is used in liquid chromatography in flash and preparative applications. By using analytical cell can be detector used also for analytical purpose.

The unit is designed as stand alone with 90–264 V power supplyand RS232 controlling. It is derived from OEM build-in unit with powering 24-36 V DC.

The unit offers many advantages:

  • wavelength setting from 200 up to 600 nm according used interference filter
  • lamp work hours are counted using the built-in counter
  • the cell is easy to replace from the side of the detector
  • powering by 90–264 V
  • unit is controlled by RS232 interface
 

Preparative cell PLCC 15 (delivered with unit):

volume/optical path(adjustable) 45 µl / 0.3 mm (basic)
55 µl / 1.4 mm
70 µl / 2.4 mm
connecting tubing with OD = 1/8", thread 1/4"-28
maximal flow rate 500 ml/min
Typical spectral half-width According used interference filter
Light source Deuterium discharge lamp
Time constant 0 s; 0.3 s; 0.6 s; 1.5 s
Output for integrator 1 V/AU (in digital form only)
Communication RS232
Power supply 90–264 V
Power input 90 VA
Dimensions W x H x D 220 x 170 x 450 mm
Weight 7.9 kg
Noise level at empty cell (254 nm, TC 1 s) ± 2 x 10-5 AU
Drift at empty cell (254 nm after 1 h) 1 x 10-3 AU/hr
Materials in contact with mobile phase PTFE; fused silica, stainless steel