SAPPHIRE 600 Ex Detector with optical connectors

UV-VIS Detector with variable wavelength 190 up to 600nm and optical connectors.

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
67SHEX0X

 Sapphire 600Ex is a detector with continuously variable wavelength in the range of 190 – 600 nm and a noise level ± 0.5 x 10-5 AU.

Cell is installed out of unit and it is connected to the unit using fiber optic cable with SMA connectors.

It is possible connect the detector to a cell for ITP or with cell PLCC 20 use it for preparative or flash chromatography.

The unit is easy to operate, reference and sample signals are available for detector diagnostics as well as information on lamp operating hours.

Output signal is available in both digital and analogue form on the output connector. It is possible to control the unit using the keypad or an RS232 interface.

The detector is equipped with a high standard deuterium and tungsten lamp in a special socket that enables easy changing.Wavelength can be programmatically changed during the performance of an analysis. The detector performs automatic wavelength calibration after the lamp has been switched on.

Cell and fibre optic cable are not include.

 

Ecom cell used with unit:

Preparative cell PLCC 20 (optionally)

Optical path – Basic continuously changeable 0.1 to 1.5 mm
Discontinuous changes of basic optical path +1 and +2 mm
Connecting thread of inlet nipple NPT 1/4”-18
Inlet nipple I.D. 1/4” (6.35 mm)
Internal cell diameter 8 mm
Maximum flow rate 15 l/min by pressure 0.5 MPa
 

 

Accuracy of adjustment ± 1 nm
Typical spectral half-width 6 nm
Reproducibility ± 0.5 nm
Light source Deuterium discharge lamp
Time constant Fast, middle and slow
Output for integrator 1 V/AU
Communication RS232
Power supply 100, 115 and 230, 50/ 60 Hz)
Power input 80 VA
Dimensions W x H x D 220 x 170 x 450 mm
Weight 11 kg
Noise level at empty cell (254 nm, TC 1 s) ± 0.5 x 10-5 AU
Drift at empty cell (254 nm after 1 h) 1 x 10-4 AU/hr
Programming 3 change of wavelength during analysis
Output and input optical connector SMA
Adjustable offset ± 50 mV