UV Detector LCD 2083.61

 Four cells, 190 up to 360 nm, build in recorder

Availability
This product cannot be purchased.
Manufacturer
ECOM spol. s r.o.
Part number
72000001

 Spectrophotometric detector LCD 2083.61 with continuously variable wavelengths in the range of 190 – 360 nm is equipped with four cell for preparative chromatography. The cells work simultaneously on the same wavelength. The unit is assembled with digital recorder for data storing. But the output signal is available in analogue and digital form too.

The detector is equipped with a high standard IST deuterium lamp in a special socket making possible an easy changing.

The unit is easy for operating, reference and sample signal is available for detector diagnostics and lamp live reading. It is possible to control it using keyboard or external signals.

 

 

Accuracy of adjustment ± 1 nm (Difference between some channel is 4 nm)
Typical spectral half-width 6 nm
Reproducibility ± 0.5 nm
Light source Deuterium discharge lamp
Time constant 1.0 sec.
Digital output RS 232
Output for integrator 1 V/AU
Power supply 230 V ± 10%, 50 Hz (115 V± 10%, 60 Hz)
Power input 80 VA
Dimensions W x H x D 220 x 320 x 470 mm
Weight 12 kg
Noise level at empty cell (254 nm, TC 1 s) ± 2.5 x 10-5 AU
Drift at empty cell (254 nm after 1 h) 5 x 10-4 AU/hr
Materials in contact with mobile phase PTFE; fused silica, stainless steel, PEEK
Wavelength 190 - 360 nm