Dual UV detector version measure at two fix wavelengths which are adjusted in production according to customer order in range of 200 – 600 nm.
Related variants
Categories
Manufacturer
ECOM spol. s r.o.
Part number
TOYL220Y
- Description
- Parameters
- Files to download 1
The unit is supplied with AMY17 PCB board of dimensions 100 x 160 mm (3.94 x 6.30 in) . It powers the lamps and control the unit.
Most important features of TOY18DAD L detectors:
•small dimensions
•built-in lamp working hours counter
•easy replacement of flow cell
•measuring at more wavelengths simultaneously
•sophisticated diagnostic software
To the AMY17 board can be optionally connected:
•Analog outputs board
•LAN and USB board
•Display with keyboard
•Leakage sensor
Number of channels (Signals) | 2 |
---|---|
Scan | 200 - 600 nm,informative |
Number of CCD elements | 256 |
Accuracy of adjustment | ± 1 nm |
Typical spectral half-width | 10 nm |
Reproducibility | ±0.5 nm |
Light source | Deuterium lamp |
Noise (Test cell, 254nm, TC 2s, 10 Hz) | ± 5 x 10-5 AU |
Drift (Test cell, 254nm) | 1 x 10-3 AU/h |
Time constant | 0.5 s, 0.75 s, 1.0 s, 2.0 s, 4.0 s, 8.0 s, 16.0 s, 0.2 s, 0.1 s |
Maximal flow rate | 500 or 3 000 ml/min depending on the cell |
Communication | RS232; optionally USB and LAN |
Power supply | 24 -36 V DC |
Power input | 48 W |
Dimensions W x H x D | 220 x 120 x 110 mm (8.67 x 4.72 x 4.33 in) |
Weight | 1.90 kg (4.19 lb) |
Materials in contact with mobile phase | FEP; fused silica, stainless steel. PEEK |
Optical cables | no |